How to use it
Use the two-page PDF to structure a development review and identify the evidence still needed for your material, process and intended use. Keep the method, sample identity and reporting basis with each result.
Key checks in the PDF
- Link the powder lot to the forming, thermal, cleaning and packaging steps actually used.
- Assess relevant phases and chemistry after processing; a nominal HA/TCP ratio alone does not establish an identical final surface.
- Report how porosity, pore size and interconnectivity were measured instead of treating them as interchangeable descriptors.
Relevant material pages